beautypg.com

JULABO LH 50 Highly Dynamic Temperature Control Systems User Manual

Page 20

background image

Operating controls and functional elements

20

2.3

Key to swap line 3 on the VFD

ID xx

Identification no. of control module RD

L xx

Capacity in %

E xxx.xx or I xxx.xx Actual value (external or internal)

P x.xx

Pump pressure in bar

2.4

Int

Key to swap line 1 on the VFD

Actual value Int or Ext alternating with line 3

2.5

Key to indicate the safety temperature on the VFD

Line 2: TANK - Safety temperature in internal bath
Line 3: RES - Safety temperature in internal reservoir

2.6

Adjustable excess temperature protection (safety temperature)
Used for setting the safety temperature in the internal bath, called
"TANK" on the display

2.7

Adjustable safety temperature.
Used for setting the safety temperature in the internal reservoir,
called "RES" on the display

2.8

Key for automatic filling and air purge of the unit as well as the
connected external system.

3 Socket

board

3.1

REG+E-PROG

Programmer input and temperature recorder output

3.2

STAND-BY

Stand-by input (for external emergency switch-off)

3.3

ALARM

Alarm output (for external alarm signal)

3.4

EXT.

Pt100

Connector for external measurement and control sensor

3.5

SERIAL

Interface RS232/RS485

3.6

RD

Connector: control cable of control module RD

This manual is related to the following products: