JULABO ME-18V Visco-Baths User Manual

Page 58

background image

Remote control

58

Command

Parameter

Response of circulator

in_mode_01 none

Selected

setpoint:

0 = SETPNT 1

1 = SETPNT 2

2 = SETPNT 3

3 = Last setpoint setting was carried out
through an external programmer

in_mode_04 none

Internal/external temperature control:

0 = Temperature control with internal sensor.

1 = Temperature control with external Pt100 sensor.

in_mode_05 none

Circulator in Stop/Start condition:

0 = Stop

1 = Start

in_mode_08 none

Adjusted control dynamics

0 = aperiodic

1 = standard

12.4. Status

messages

Status messages

Description

00 MANUAL STOP

Circulator in „OFF“ state.

01 MANUAL START

Circulator in keypad control mode.

02 REMOTE STOP

Circulator in „r OFF“ state.

03 REMOTE START

Circulator in remote control mode.


12.5. Error

messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-02 REFRIGERATOR ALARM

Control cable of the refrigerated circulator or MVS solenoid
valve controller short-circuited or interrupted.

-03 EXCESS TEMPERATURE

WARNING

High temperature warning.

-04 LOW TEMPERATURE

WARNING

Low temperature warning.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or interrupted.

-06 SENSOR DIFFERENCE ALARM Sensor difference alarm.

Working temperature and safety sensors report a temperature
difference of more than 35 K.