Preparations
18
5.3.
Filling / draining
Notice:
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit
on for a short moment, for example).
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit, or before moving the unit.
Store and dispose the used bath fluid according to the laws for environmental
protection.
Plexiglas and Makrolon baths are not resistent to solvents.
Avoid contamination of the bath fluid with solvents.
ME-31A
ME-16G
Filling
Take care that no liquid enters the interior of the
circulator.
The filling height is dependent upon the viscometer
used.
Please follow the manufacturer’s instructions.
After reaching the set temperature, please check
and if necessary adjust the filling height.
Recommended maximum filling level with water as
bath liquid: 25 mm below the top panel.
Recommended maximum filling level with bath oils:
30 mm below the top panel.
Draining
Press the mains switch to turn the circulator off.
Prepare a suitable bath tank to take the used bath
liquid.
ME-31A and ME-16G
Empty the bath tank with a siphon or an external
pump.