beautypg.com

JULABO ME-18V Visco-Baths User Manual

Page 50

background image

Menu functions

50

BAND HIGH / LOW – Band limitation

The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.

Setting range: 0 °C ... 200 °C

BAND HIGH

and BAND LOW allow for the limitation of the difference

between the temperatures in the internal bath and the external system to
any maximum value for the heat-up and cool-down phase.
During the heat-up phase this difference value is always added to the
actual external temperature. During the cool-down phase the difference
value is subtracted.

25

50

10 20 30 40 50 60 10 20 30 40 50 60 10 [min]

75

100

150

200

°C

t

System extern

Bad intern

BandOben

BandUnten

Sense of a band limitation:

Protection of objects and samples by
gentle temperature control

Protection of e.g. glass reactors from
thermal shock.




This manual is related to the following products: