beautypg.com

Setting the pump pressure – JULABO SL-26 Heating Circulators User Manual

Page 41

background image

HL / SL

41

9.3. Setting the pump pressure


The capacity of the circulating pump is set by adjusting the motor speed

Press the

key if a parameter is to be retained. (prior to OK)

Level 2

Submenu >Pump<

The set pump stage is displayed. Example: 1

 Confirm

by

pressing

.

 Set required pump stage by pressing

and

.

Example: 3


Example:
Adjusted Effective

Adjustable pump capacity

stage 1 ... 4

Illuminated display:

for pump pressure

Indication of the adjusted pump pressure stage in the –OFF- mode.

Indication of the effective pump pressure stage (rotational speed)
after start.

To protect the pump motor, the rotational speed changes with the load
applied, for example, depending on the viscosity of the bath fluid at
different working temperatures.

Flow rate:

22 ... 26 l/min

Pump

capacity

stage

1 2 3 4

Pump

pressure [bar]

0.4 0.5 0.6 0.7

Suction

pump

[bar]

0.2 0.26 0.33 0.4

Total

capacity

[bar]

0.6 0.76 0.93 1.1

in a loop circuit

This manual is related to the following products: