9.1.5.
Interruption after a power failure
It is possible to operate the programmer safely in case of a power 
failure. 
Important: 
For this, >Autostart< needs to be set in the >Configuration< menu. 
If the AUTOSTART function is activated, the programmer starts 
again at a point approx. 20 seconds before the interruption took 
place. However, an uncontrolled change of the bath temperature has 
occurred. 
Power failure when the instrument is on hold if start is effected via 
the built-in timer: 
If the starting time is not yet reached, the programmer is put on hold 
again. 
If the starting time is exceeded, the programmer starts immediately. 
 
 
9.1.6.
Termination of a program
A program can be terminated by pressing
. The programmer
switches back to the Start menu. 
 
Further procedures are defined in the MENU >START PROGRAM<, 
menu item >Program End <. 
 
Example: >Program End - StdBy< 
The VFD COMFORT-DISPLAY displays “OFF”.
 
 
 
Interruption after a power failure
In case of a power failure, the reaction of the circulator is the same as 
when switched off and on again with the mains switch. 