Program administration, creation – JULABO SL-26 Heating Circulators User Manual
Page 37

HL / SL
37
9.2. Program administration, creation
The integrated programmer permits fast and easy programming of 
setpoint temperature sequences. This temperature sequence is called 
program. A program is composed of individual sections (Step). 
The sections are defined by duration (t) or Gradient (°/t) and target 
temperature. The target temperature is the setpoint (SP), which is 
achieved at the end of a section. The programmer calculates the 
temperature ramp from the difference in time and temperature. 
 
Create programs
A flashing segment indicates that a number needs to be entered. 
Under the submenu “Edit”, enter a program number. 
Six programs may be stored (numbers 0 to 5). 
When the program is running, only sections with complete information for 
the target temperature and time period are considered. 
It makes sense to skip section numbers in the profile in order to use them 
later for corrections to the profile. 
Important: 
If a time of 00:00 is set for a profile, the profile is continued with the next 
section only after the setpoint temperature (±0.2 °C) is reached. 
Example: 
Step 
(Nr.)
1 3 5 6 9 11
14
Setpoint (°C)
50 50 100
100
80 80 20
Time
(h:m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30