Preparations 5.1. installation, Bath fluids – JULABO SL-26 Heating Circulators User Manual
Page 18
Preparations
18
5. Preparations
5.1. Installation
Heating circulators
Place the unit on an even surface on a base made of nonflammable
material.
5.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of
65 °C is used, only supervised operation is possible.
Water: The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.