JULABO SL-26 Heating Circulators User Manual
Page 28

Safety installations, warning functions
28
OverTemp 87.0°C
SubTemp 83.0°C
Setpoint3 85.0°C
The warning functions are only triggered when the actual bath 
temperature, after start from the „OFF“ or „rOFF“ mode, lies within 
the set limits for 3 seconds. 
 
 
 
 
 
 
 
 
 
 
 
 Setting
>WARNING<
A mere warning function with optical and audible warning signal 
(interval tone) A message appears on the VFD COMFORT-
DISPLAY: 
or
OverTemp SubTemp 
 
The LCD –DISPLAY shows the message: 
 Setting
>ALARM<
Temperature limit with shutdown of heater and circulating pump. 
An audible alarm sounds (continuous tone) and a message appears 
on the VFD COMFORT-DISPLAY: 
or
OverTemp SubTemp 
 
The LCD -DISPLAY shows the message 
Press
for help on the LCD