Safety installations, warning functions

28

OverTemp 87.0°C

SubTemp 83.0°C

Setpoint3 85.0°C

The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within
the set limits for 3 seconds.










 Setting

>WARNING<

A mere warning function with optical and audible warning signal
(interval tone) A message appears on the VFD COMFORT-
DISPLAY:

or

OverTemp SubTemp

The LCD –DISPLAY shows the message:

 Setting

>ALARM<

Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message appears
on the VFD COMFORT-DISPLAY:

or

OverTemp SubTemp

The LCD -DISPLAY shows the message

Press

for help on the LCD