To maintain the full cooling
performance, clean the condenser (C)
from time to time.
1. Switch off device by pressing the
main power switch and
2. disconnect mains cable from power
source.
3. Remove 2 screws
4. Lift cover upwards.
5. Remove dirt at condenser by
suction cleaning.
6. Close cover and
7. Fix by means of screws.
8. Unit is ready for operation.
Cleaning:
Clean the outside of the unit using a wet cloth and low surface tension water.
The Cryo-Compact Circulator is designed for continuous operation under
normal conditions. Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended by JULABO. To
avoid contamination, it is essential to change the bath fluid from time to time.
Repairs:
Before asking for a service technician or returning a JULABO instrument for
repair, please contact an authorized JULABO service station.
When returning the unit:
Clean the unit in order to avoid any harm to the service personnel
Attach a short fault description.
When returning a unit, take care of careful and adequate packing.
JULABO is not responsible for damages that might occur from insufficient
packing.
JULABO reserves the right to carry out technical modifications along with
repairs to provide improved performance of a unit.