JULABO CF40 Cryo-Compact Circulators User Manual
Page 23

Cryo-Compact Circulator
23
6.5.
Setting the temperatures
Setting can be carried out in the start/stop condition.
1.
Press one of the keys
for a short moment.
The setpoint value instead of the actual value is indicated on the
display for about 8 seconds.
The value can now be changed.
2.
Change value:
Press
to set a higher value.
Press
to set a lower value.
Keep the keys depressed for the value to change fast.
3.
Press enter
to store the value.
ESC
Press ESC to update the display immediately, or the unit
automatically returns to the effective display after
about 30 seconds
.
6.6.
AUTOSTART ON / OFF
The Cryo-Compact Circulator has been configured and supplied by
JULABO according to N.A.M.U.R. recommendations. This means for
the start mode, that the unit must enter a safe operating state after a
power failure (non-automatic start mode). This safe operating state is
indicated by „OFF“ on the LED temperature display. A complete
shutdown of the main functional elements such as compressor and
circulating pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the
start of the Cryo-compact circulator directly by pressing the mains
power switch or using a timer.
1.
Keep depressed enter
and
2.
turn on the unit with the mains power switch.
For a short while the LED DISPLAY indicates the effective start
mode:
AUTOSTART on.
AUTOSTART off.