Operating procedures
20
6. Operating
procedures
6.1. Power
connection
Caution:
Only connect the unit to a power socket with an earthing contact (PE –
protective earth)!
The power supply plug serves as a safe disconnecting device from the line
and must always be easily accessible.
Never operate the unit with a damaged mains power cable.
Regularly check the mains power cables for damage.
We disclaim all liability for damage caused by incorrect line voltages!
Make sure that the line voltage and frequency match the supply
voltage specified on the type plate.
6.2. Filling
Notice:
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Guideline:
A volume change of 12 % per 100 °C temperature variation is to be
considered.
Take care that no liquid enters the interior of the Cryo-
compact circulator.
Connect the tubing from the external system to the pump
connectors and check for leaks.
Check to make sure that the drain tap (7) is closed.
Recommendation:
For filling, use for example an measuring jug with nuzzle.
Recommended maximum filling level with water as bath fluid:
30 mm below the tank rim
Recommended maximum filling level with bath oils:
40 mm below the tank rim