Operating procedures

20

6. Operating

procedures

6.1. Power

connection

Caution:

 Only connect the unit to a power socket with an earthing contact (PE –

protective earth)!

 The power supply plug serves as a safe disconnecting device from the line

and must always be easily accessible.

 Never operate the unit with a damaged mains power cable.

 Regularly check the mains power cables for damage.
 We disclaim all liability for damage caused by incorrect line voltages!



Make sure that the line voltage and frequency match the supply
voltage specified on the type plate.



6.2. Filling

Notice:

Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.

Guideline:
A volume change of 12 % per 100 °C temperature variation is to be
considered.

Take care that no liquid enters the interior of the Cryo-
compact circulator.

Connect the tubing from the external system to the pump
connectors and check for leaks.

Check to make sure that the drain tap (7) is closed.

Recommendation:
For filling, use for example an measuring jug with nuzzle.

Recommended maximum filling level with water as bath fluid:
30 mm below the tank rim

Recommended maximum filling level with bath oils:
40 mm below the tank rim