beautypg.com

JULABO CF40 Cryo-Compact Circulators User Manual

Page 21

background image

Cryo-Compact Circulator

21

 Turn the mains switch (1) on

(Switching on - see page 22)

 Switch on unit. To do so press button

for approx. 4 seconds.

 Tempering fluid is pumped into the externally connected system.

Refill fluid.

 The Cryo-Compact Circulator is ready for operation.

Important:

When using a bath fluid, the change in volume in case of change
in temperature has to be respected. Fill in a little amount of bath
fluid only so that the low level alarm is not triggered.

Low level alarm is triggered at the following liquid level:

CF30 75 to 80 mm below the tank rim

CF40 80 to 85 mm below the tank rim

When reaching the working temperature, check the liquid level. If
the cooling coil is not completely covered with bath fluid, refill it.

This manual is related to the following products: