Cryo-Compact Circulator

31

in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of recirculating cooler

version none

Number

of

software version (V X.xx)

status

none

Status message, error message (see page 31 )

in_pv_00

none

Actual bath temperature.

in_pv_01

none

Heating power being used (%).

in_pv_03

none

Temperature value registered by the safety sensor.

in_pv_04 none

Setpoint

temperature of the excess temperature

protection

in_sp_00 none

Working temperature

in_mode_05

none

Cryo-Compact Circulator in Stop/Start condition:
0 = Stop
1 = Start

10.4. Status messages

Status messages

Description

00 MANUAL STOP

Cryo-compact circulator in „OFF“ state.

01 MANUAL START

Cryo-compact circulator in keypad control mode.

02 REMOTE STOP

Cryo-compact circulator in „r OFF“ state.

03 REMOTE START

Cryo-compact circulator in remote control mode.

10.5. Error

messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or
interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 35 K.

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.