Description

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE

MEASUREMENT ALARM

Error in A/D converter.

-14 EXCESS TEMPERATURE

PROTECTOR ALARM

Excess temperature protection alarm

-20 WARNING: CLEAN

CONDENSOR OR CHECK
COOLING WATER CIRCUIT OF
REFRIGERATOR

Cooling of the condenser is affected.
Clean air-cooled condenser.

-21 WARNING: COMPRESSOR

STAGE 1 DOES NOT WORK

Compressor does not work.

-33 SAFETY SENSOR ALARM

Excess temperature sensor short-circuited or
interrupted.