5.2. Temperature

application to external, closed systems

The Cryo-Compact Circulator is used for temperature application to external,
closed systems (loop circuit)

Caution:

Securely attach all tubing to prevent slipping.

Notice:

Flood hazard!

If the liquid levels in the Cryo-Compact Circulator bath and the external system
are at different heights, overflowing must be prevented after the power has
been turned off.


SERI AL

T16A

T16A

 Unscrew the M16x1 collar nuts on the

pump connectors with a 19 mm (3/4“)
wrench and remove the sealing disks.
Using the collar nuts, screw on the tubing
connection fittings (for tubing 8 mm or 12
mm in diameter) delivered with the unit and
tighten firmly. (Pressure pump: 8, Return:
9)

 Push on the tubings, and secure with tube

clamps.

 Attach the tubing to the connectors of the

external closed system, e.g., an instrument
with a pressure-resistant temperature
jacket or a temperature coil, and fasten
with tube clamps to prevent slipping.

Tubing see page 18

Return flow safety device

For this reason, shut-off valves can be integrated in the loop circuit.

Order No.

Description

8 970 456

Shut-off valve (suitable up to +90 °C)

8 970 457

Shut-off valve (suitable up to +200 °C)