JULABO CF40 Cryo-Compact Circulators User Manual
Page 17
Cryo-Compact Circulator
17
5.2. Temperature
application to external, closed systems
The Cryo-Compact Circulator is used for temperature application to external,
closed systems (loop circuit)
Caution:
Securely attach all tubing to prevent slipping.
Notice:
Flood hazard!
If the liquid levels in the Cryo-Compact Circulator bath and the external system
are at different heights, overflowing must be prevented after the power has
been turned off.
SERI AL
T16A
T16A
Unscrew the M16x1 collar nuts on the
pump connectors with a 19 mm (3/4“)
wrench and remove the sealing disks.
Using the collar nuts, screw on the tubing
connection fittings (for tubing 8 mm or 12
mm in diameter) delivered with the unit and
tighten firmly. (Pressure pump: 8, Return:
9)
Push on the tubings, and secure with tube
clamps.
Attach the tubing to the connectors of the
external closed system, e.g., an instrument
with a pressure-resistant temperature
jacket or a temperature coil, and fasten
with tube clamps to prevent slipping.
Tubing see page 18
Return flow safety device
For this reason, shut-off valves can be integrated in the loop circuit.
Order No.
Description
8 970 456
Shut-off valve (suitable up to +90 °C)
8 970 457
Shut-off valve (suitable up to +200 °C)