JULABO CF40 Cryo-Compact Circulators User Manual
Page 24
Operating procedures
24
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid
any hazardous situation to persons or property.
The Cryo-compact circulator does no longer conform to N.A.M.U.R.
recommendations.
6.7.
Remote control: activate – deactivate
(Interface OFF)
(Interface On)
The Cryo-Compact Circulator is to be prepared for remote control
by a personal computer via the serial interface RS232.
Set the interface item
from >IOFF<
(Interface OFF)
to
>ION<
(Interface On).
Remote control: activate – deactivate:
Switch off the Cryo-Compact Circulator by pressing the mains
switch and wait approx. 5 seconds.
Keep depressed the keys
and enter
simultaneously
and turn on the unit with the mains power switch.
>I OFF< No remote control via RS232 (Factory setting)
>I On<
Remote control via RS232
The software version and the type of unit is indicated
(see example on the left).
The display "r OFF" indicates the unit is ready to be operated
via remote control.