Operator responsibility – Safety instructions

10

Warning functions and safety installations

Excess temperature protection

adjustable from 0 °C ... 220 °C

Low liquid level protection

float switch

Classification according to DIN 12876-1

class III

Alarm message

optical + audible (permanent)

Warning message

optical + audible (in intervals)

Overload protection

for compressor and pump motor

Supervision of working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >35 K






Environmental conditions according to IEC 61 010-1:

Use indoors only.

Altitude up to 2000 m - normal zero.

Ambient

temperature:

see

Technical specifications

Humidity:

Max. relative humidity 80% for temperatures up to +31 °C,

linear decrease down to 50% relative humidity at a temperature of +40 °C

Max. mains voltage fluctuations of ±10% are permissible.


Protection class according to IEC 60 529

IP21

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not suitable for use in explosive atmosphere


EMC requirements according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes),
Class A (industrial and commercial range).