JULABO CF40 Cryo-Compact Circulators User Manual
Page 10
Operator responsibility – Safety instructions
10
Warning functions and safety installations
Excess temperature protection
adjustable from 0 °C ... 220 °C
Low liquid level protection
float switch
Classification according to DIN 12876-1
class III
Alarm message
optical + audible (permanent)
Warning message
optical + audible (in intervals)
Overload protection
for compressor and pump motor
Supervision of working sensor
plausibility control
Reciprocal sensor monitoring between
working and safety sensors
difference >35 K
Environmental conditions according to IEC 61 010-1:
Use indoors only.
Altitude up to 2000 m - normal zero.
Ambient
temperature:
see
Technical specifications
Humidity:
Max. relative humidity 80% for temperatures up to +31 °C,
linear decrease down to 50% relative humidity at a temperature of +40 °C
Max. mains voltage fluctuations of ±10% are permissible.
Protection class according to IEC 60 529
IP21
The unit corresponds to Class I
Overvoltage category
II
Pollution degree
2
Caution:
The unit is not suitable for use in explosive atmosphere
EMC requirements according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes),
Class A (industrial and commercial range).