Rs-3000 – INFICON RS-3000 Power Supply Controller User Manual
Page 17
RS-3000
Power Supply Controller
instruments
Table 3-3
Back-panel 15-pin D-sub Female Pin Descriptions
PIN
NUMBER
DESCRIPTION of Connector Labeled External Control
I/O
Direction
1
Analog ground of the RS-3000 presented to remote controller
(used with Voltage control [pin #9])
ground
pass thru
2
Along with pin 10, this line presents the RS-3000 digital ground to
the remote controller
ground
pass thru
3
Interlock status indication: grounded (to pin 2) or digital low =
satisfied, open or digital high = interlock error (indicated by front
panel LED)
Input
w/ pull-up
4
Process abort (digital low = process abort)
Input
w/ pull-up
5
Process START event (digital falling edge = START event)
Input
w/ pull-up
6
Fault reset (digital falling edge = reset event)
Input
w/ pull-up
7
Remote out (digital high = remote enabled)
Output
w/ pull-up
8
Ready (digital high = ready enabled)
Output
w/ pull-up
9
Voltage control (remote voltage control, via RS-3000, of the
T3000)
Input
10
Along with pin 2, this line presents the RS-3000 digital ground to
the remote controller
ground
pass thru
11
Process timer reset (digital falling edge = reset event)
Input
w/ pull-up
12
Process STOP event (digital falling edge = STOP event)
Input
w/ pull-up
13
Remote/Local START mode
Input
w/ pull-up
14
Busy (digital high = active)
Output
w/ pull-up
15
Fault (digital high = active)
Output
w/ pull-up
Specifics
Page 3 - 6
Section 3