E

Attrition (PTFE particles) at bearing

o Tools were operated without or with too little liquid; clean tool, if necessary

exchange bearing. Please note the filling quantities of test tubes (p. 23).

If malfunctioning occurs that cannot be corrected with the above instructions, please
inform your authorized Heidolph Instruments dealer immediately.

TECHNICAL DATA

SilentCrusher S

Characteristics

Values

Connection Voltage

100

– 240 V / 47 - 63 Hz

Power Consumption (Rated Output)

24 W

RPMs

15.000

– 75.000 rpm (referred to water)

Operating Mode

Short-term operation, 1.5 minutes, automatic
shut-off, red control light indicates shut-off.
Reset by removing dispersing tool.

Drive Protection

Overheating protection switches drive unit off
in case of overheating; blinking red control
light indicates this condition. Reset by
removing dispersing tool.

Control Light for “Device On”

Control light (green) on power supply

Dimensions (in mm) Drive Unit

Width 125; height 197; depth 125

Dimensions (in mm) Power Supply

Width 106; height 30; depth 65

Total weight

1.1 kg

Operating Temperature

0

– 40 °C

Storage Temperature

- 20 to + 80 °C

Relative Humidity

85 % (without condensation)

Type of Protection according to EN
60529

IP 30

Safety Class according to VDE 0100