Cleaning and maintenance, Drive unit, Dispersing tool – Heidolph SilentCrusher S User Manual

Page 30: Cleaning, Disassembly and bearing replacement

background image

30

E

After cooling (about 5 min.), this control light will go out and dispersion can be continued.

When the PTFE bearing in the dispersion tool becomes worn, it must be

replaced. See the Cleaning and Maintenance chapter.

Wear can be recognized from the increased play between the rotor (O) and the

PTFE bearing (Q) or the increased level of noise.

Caution: The tool and fluid for dispersion are heated by the energy input.

CLEANING AND MAINTENANCE

5

Drive Unit

For cleaning, you can wipe down the housing and surface of the device with a moist cloth
(mild soap solution).

Note:

Never use chlorine bleach, chlorine-forming cleansers, abrasives, ammonia,

steel wool or cleaning agents with metal particles. They can damage the surface

of the device.

The device is maintenance-free. Any repairs required must be performed by a Heidolph
Instruments authorized technician. Contact your Heidolph Instruments dealer or a
representative of Heidolph Instruments (see page 39).

6

Dispersing Tool

To avoid contamination, the dispersing tool must be cleaned after every

dispersing process.

6.1

Cleaning

Clean the dispersion tool with a suitable cleaning fluid (depending on the medium
processed). Run the dispersing tool in the cleaning fluid so that the internal parts are also
cleaned. If the tool is extremely dirty, the dispersing tool can be disassembled for cleaning.