Behaviour after power cut

After interruptions such as power failures or deliberate disconnection (e.g. timer) of the AC

outlet, the magnetic stirrer remains in stand-by-mode (no heating, no stirring). This

function is in operation when lower right of display shows “o”.

11

Residual heat display

To reduce the risk of burns, MR Hei-End

displays the residual heat of the hotplate.

Any time the hotplate has a temperature

higher than 50°C after the rocker switch (A)

has been switched off, the warning shown

here will be displayed. The instrument will

switch off fully once the temperature has dropped below 50°C.

12

Connecting the interfaces

Switch the instrument off before plugging in

the interface cable.
Please use an interface cable with SUB-HDD

15 connectors and connect it to the socket

(K) on the rear panel of the instrument.

12.1 RS232 interface

Use the RS232 interface and a suitable

program (e.g. Microsoft

©

Hyperterminal), or a

suitably programmed interface to enter and

read out the parameters described in 12.1.1. Configure Interface to receive data in 7 bit

ASCII format. Sending and receiving data has to end with a line feed.