1.5 Operating conditions
➨
The pumps have no approval for operation in or for pumping of potentially
explosive atmospheres.
➨
The pumps are
not suitable to pump
- unstable substances and substances which react explosively under
impact (mechanical stress) and/or when being exposed to elevated
temperatures without air,
- self inflammable substances,
- substances which are inflammable without air and
- explosive substances.
• The pumps are
not suitable for pumping substances which may form deposits
inside the pump. Deposits and condensate in the pump may lead to increased
temperatures even to the point of excessing the maximum permitted tempera-
tures!
• If there is a danger of formation of
deposits in the pump chamber (check inlet
and outlet of the pump), inspect the pump chambers regularly and clean if neces-
sary.
• The pumps are
not suitable for pumping dust and have no approval for opera-
tion below ground.
If pumping
different substances, it is recommended to purge the pump with air or
inert gas prior to changing the pumped media in order to pump out residues and to
avoid reactions of the pumped substances with each other and with the pump ma-
terials.
Take into consideration interactions and chemical reactions of the pumped
media. Ensure that the materials of the wetted parts are compatible with the pumped
substances, see section ”Technical data”.
1.6 Safety during operation
➨
Adopt suitable measures to prevent the release of dangerous, toxic, explosive,
corrosive, noxious or polluting fluids, vapors and gases. In case install an ap-
propriate collecting and disposal system and take protective action for pump and
environment.
➨
Prevent any part of the human body from coming into contact with vacuum.
➨
The user must take suitable precautions to prevent any formation of explosive
mixtures in the expansion chamber or at the outlet. In case of e.g. a diaphragm
crack, mechanically generated sparks, hot surfaces or static electricity may ignite
these mixtures. Use inert gas for gas ballast or venting if necessary.
➨
Potentially explosive mixtures at the outlet of the pump have to be drained ap-
propriately, sucked off or diluted with inert gas to non-explosive mixtures.
☞
Pay attention to the symbol ”hot surfaces” on the equipment. Adopt suitable
measures to prevent any danger arising from the formation of hot surfaces or
electric sparks. Provide a suitable protection against contact if necessary.
☞
Pumping at
high inlet pressure may lead to overpressure at the gas ballast
valve. Pumped gases or condensate might be pushed out in case the valve is
open. If an inert gas supply is connected, ensure that the inlet pipeline is not con-
taminated.
NOTICE