7 maintenance and repair, Notice – Heidolph ROTAVAC vario control User Manual
Page 9
![background image](/manuals/561945/9/background.png)
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999223 / 08/07/2009
page 9 of 34
1.7 Maintenance and repair
Wear parts have to be replaced regularly. In case of normal wear the lifetime of
the diaphragms and valves is > 10000 operating hours. Bearings have a typical du-
rability of 40000 h (see section 7
”Replacing diaphragms and valves”).
➨
Isolate equipment from mains and wait two minutes before starting mainte-
nance to allow the capacitors to discharge.
☞
Ensure that the pump cannot be operated accidentally. Never operate the pump
if covers or other parts of the pump are disassembled. Never operate a defective
or damaged pump.
☞
Attention: The pump might be contaminated with process chemicals which have
been pumped during operation. Ensure that the pump is decontaminated before
maintenance and take adequate precautions to protect people from the effects of
dangerous substances if contamination has occurred.
• Before starting maintenance vent the pump, isolate the pump and other compo-
nents from the vacuum system. Allow sufficient cooling of the pump. Drain con-
densate, if applicable.
Ensure that
maintenance is done only by suitably trained and supervised techni-
cians. Ensure that the maintenance technician is familiar with the safety procedures
which relate to the products processed by the pumping system.
In order to comply with law (occupational, health and safety regulations, safety at
work law and regulations for environmental protection) vacuum pumps, components
and measuring instruments returned to the manufacturer can be repaired only when
certain procedures (see section 6
”Questions / Repair”) are followed.
NOTICE