Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999223 / 08/07/2009

page 8 of 34

• Comply with applicable regulations when disposing of chemicals. Take into con-

sideration that chemicals may be polluted.

Take adequate precautions to protect people from the effects of dangerous sub-

stances (chemicals, thermal decomposition products of fluoroelastomers), wear

appropriate safety-clothing and safety glasses.

• Use only

genuine spare parts and accessories. Otherwise safety and perform-

ance of the equipment as well as the electromagnetic compatibility of the equip-

ment might be reduced.

Possibly the CE mark becomes void if not using genuine spare parts.

• Failure of the pump (e.g. due to power failure) or of connected components,

parts of the supply or change of parameters must not lead to a critical dangerous

situation under any circumstances. In case of diaphragm cracks or leaks in the

manifold pumped substances might be released into the environment or into the

pump housing or motor. Comply especially with notes on operation and use and

maintenance.

• Due to the residual

leak rate of the equipment, there might be an exchange of

gas, albeit extremely slight, between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of the pumped substances or

the environment.

Check liquid level in the catchpot of the exhaust waste vapor condenser (accessory)

regularly and drain condensate in time.

Do not start the pump if the pressure difference between inlet and outlet port ex-

ceeds 1.1 bar at maximum.

Prevent any backpressure of gases and the backflow of condensates.

Never suck liquids or dust into the pump.

Provide appropriate protective measures (i.e precautions which allow for the require-

ments of the respective application) even for the case of failure and

malfunction.

Failure of the pump (e.g. due to power failure) or of connected components, of parts

of the supply or change of parameters must not lead to a critical dangerous situation

under any circumstances.

The motor is protected by a

temperature sensor at the circuit board: Current

limitation if the temperature at the circuit board is higher than 70°C, switching off the

pump if the temperature is higher than 85°C. In case of blockade of the motor (after

10 attempts to start-up) the pump is switched off.

If the pump is switched off due to safety measures, manual reset is necessary. Isolate

the pump from mains. Eliminate the cause of failure before restarting the pump.

The A-weighted emission sound pressure level of the pump does not exceed 70

dB(A). Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 with

standard silencer or exhaust tube at outlet.