10.12 The unit does not liberate any known poisonous substances. Caution is recommended as any
exposed surface may have acid vapor residue due to usage. to maintain the maximum life of the
electrical components, the inert gas line must remain connected and operational even when the unit is
not in service.
10.12.1 – waste from the drain line should be neutralized and disposed of with a waste
management service.
10.12.2 – any liquid discharge should be treated as potentially acidic and disposed of using a
waste management service.
10.13 All surfaces and accessories that are exposed to elevated temperatures within the digestion
chamber are considered a burn risk and should be treated with caution.
10.14 If this equipment is used in a matter other than specified, the protection provided by the
equipment may be impaired.
10.15 Do not replace power supply cords with inadequately rated replacement cords.
It is recommended that these are supplied through environmental express.
10.16 All replacement parts are needed to be supplied by Environmental Express.
This includes the replacement of valves, tubing, heater elements, and control boards.
10.17 Replacement FuSES are rated for a maximum voltage of 250v.
Each individual fuse is rated as 6A or 12A depending on the system purchased.
10.18 When servicing equipment, service personnel should:
10.18.1 – Remove power supply when working in the electrical cabinet.
10.18.2 – Prolonged use of the unit may result in acid residue on the various surface. The unit
should be cleaned periodically and hand protection should be worn when servicing the unit.
10.18.3 – All personal protective gear should be work when using equipment. This includes
safety eyewear, gloves, and closed toed footwear.
10.18.4 – Servicing of the equipment should be performed in a well ventilated area