DIGIWEIGH DWP-C2X User Manual
Page 13

10
The wire of transfer
is
broken
or
damaged
Press at different area of the
transfer wire, the data changes
obviously
Reconnect the wires(protect
from short-circuit
Serious scale body
moisture
See vapor inside after open the
front cover
Dry wet parts
Insufficient
welding
during maintenance
Welding points not smooth or with
small gaps
Redo the welding
Insert loose
check the loose by eyes, or shake
the insert, the value changes
Re plug or replace insert
Not
show
"0"
when the scale
empty
The preheat time is
not enough when
startup
Preheat time 3-5 minutes
Shows "0" after 3-5 minutes
Startup
when
the
scale is not hung up
Should startup with hung up
empty scale
Zero
Huge tolerance of
weighing value
Inaccurate reference
standard
Reference to "standard" is not a
standard weight, not test or test
for a long time
Calibrate with standard weights
or weight together with a new
scale from the distributor.
Inaccurate
placement
of
the
hanger
Check if any rope messed up,
over side load, protruding after
knotted during weighing
Avoid any
slapping. No
centered, make sure no
stuck of the hanger
Remote
not
working
Low battery
Check the battery capacity
Replace battery
Long distance
Move closer to the scale
Button failure
Other buttons work normally
Replace the remote