DIGIWEIGH DWP-C2X User Manual
Page 10

7
11
Press(▲)
Press [*]
[ brt *]
[ brt 6]
Set backlight brightness for LCD
Press(▲) to select the brightness:
Brt=1~64
Press [
0
] on remote, each press
minus the parameter by 1
Quit the setting and back to weighing
status
Remarks:
1.Can quit "Setup" at any time and return to weighing status, the data becomes effective only after all
processes completed.
2. The bigger setup data the more stable weighing
,and with longer time for staying stable. And vice
versa.
3. The original setup gravity is 9.7936. The user can adjust it into local one; program will
automatically correct the gravity. If the user has calibrated at the local, the gravity doesn
’t
need to be adjusted.
Chapter 8 Weighing Demarcation & Compensation Demarcation
1. Calibration (this operation must be after password setting. The initial password is 33333)
Press (▲) for plus, and (→0←) for minus, (*) for confirmation and enter into next step, press (H) to
quit setting mode.
Ste
p
Operation
Display
Remarks
1
Press (*)
Press(▲)
Press(*)
[ CAL 1 ]
[ CAL 2 ]
Option: CAL 1 Parameter setting CAL 2 Calibration
Switch to calibration operation
Enter into calibration status
2
[ *****]
[ 0 0 0 0 0 ]
check scale ZERO
Make sure the scale is empty and check if the display is
zero. If not, press [TARE ON/OFF] on the scale panel or
(
0
) on the remote to make it zero. If it is zero, enter to next
step.
3
Load
weight
Press (*)
[ ##### ]
Load weight( load weight as close as to the max. capacity
Load weight on the scale, #### is the weight after
loading.
Enter into input of calibrating weight status