7 calibration – DIGIWEIGH DWP-PJ-P User Manual
Page 9
![background image](/manuals/557891/9/background.png)
DW-PJ Pallet Jack Scale
- 8 -
Press【↑】
〖002000〗
〖003000〗
For example: F·S=3000
5
Press【﹡】
Press【→】
〖FLt 10〗
〖FLt 00〗
Sets digit Filter parameters :00-99
The display will updata faster and filter faster as
The filter parameter is changed from 99-00
For example: FLt = 00
6
Press【﹡】
Press【→】
Press【→】
Press【↑】
〖AUtP00〗
〖AUtP00〗
〖AUtP00〗
〖AUtP10〗
Sets Automatic Power Off function
AUtP=00 Not automatic power off
AUtP=01 Automatic power off
th digit express the choice of zero trace range (1--9):
1:0.4 d 2:0.8 d 3:1.2 d 4:1.6 d 5:2d
6:2.4 d 7:2.8 d 8:3.2 d 9:3.6 d
decimal digit express the choice of zero set
decimal digit =0 no zero set at start operation
decimal digit ≧1 zero set at start operation 20%FS
For example:AUtP=10
(AUtP=10 when leaving the factory)
7
Press【﹡】
〖Adr 00〗
continuous output :Adr=00
8
Press【﹡】
Press【↑】
〖b 2400〗
〖b 4800〗
Baud rate range:1200→2400→4800→9600
For example:b=4800
9
Press【﹡】
〖 1000〗
No :1 setpoint output
Weight<1000 display LO
☆
10
Press【﹡】
〖 2000〗
No :2 setpoint output
1000< Weight<2000 display OK
Weight>2000 display HI
☆
11 Press【﹡】
〖 0〗
Press to confirm configurations and go to calibration
menu
Note:one time accumulation is allowed for weighing once. Following accumulation is
allowed for weighing only when displayed value is below 20 d.
☆ Optional LO
—OK—HI and two setpoint output PCB
☆ Press
【→】key for two seconds, LCD display for LED are turned on(A1GB-L)
7 Calibration
Calibration should be done after setting the parameters
Step
Operation
Displaying
Comments
1
Press【﹡】
Press【→】
〖CAL SP〗
〖CAL 00〗
Enters calibration
Zero the scale
2
Press【﹡】
〖------〗
〖 3000〗
Starts zero calibration and wait for calibration to
complete
3
Load
the
standard
weight for F·S on the
platform press【﹡】
〖------〗
〖 3000〗
Starts calibration and wait for calibration to
complete