7 calibration – DIGIWEIGH DWP-PJ-P User Manual

Page 9

background image


DW-PJ Pallet Jack Scale

- 8 -

Press【↑】

〖002000〗

〖003000〗

For example: F·S=3000

5

Press【﹡】

Press【→】

〖FLt 10〗

〖FLt 00〗

Sets digit Filter parameters :00-99

The display will updata faster and filter faster as

The filter parameter is changed from 99-00

For example: FLt = 00

6

Press【﹡】

Press【→】

Press【→】

Press【↑】

〖AUtP00〗

〖AUtP00〗

〖AUtP00〗

〖AUtP10〗

Sets Automatic Power Off function

AUtP=00 Not automatic power off

AUtP=01 Automatic power off

th digit express the choice of zero trace range (1--9):

1:0.4 d 2:0.8 d 3:1.2 d 4:1.6 d 5:2d

6:2.4 d 7:2.8 d 8:3.2 d 9:3.6 d

decimal digit express the choice of zero set

decimal digit =0 no zero set at start operation

decimal digit ≧1 zero set at start operation 20%FS

For example:AUtP=10

(AUtP=10 when leaving the factory)

7

Press【﹡】

〖Adr 00〗

continuous output :Adr=00

8

Press【﹡】

Press【↑】

〖b 2400〗

〖b 4800〗

Baud rate range:1200→2400→4800→9600

For example:b=4800

9

Press【﹡】

〖 1000〗

No :1 setpoint output

Weight<1000 display LO

10

Press【﹡】

〖 2000〗

No :2 setpoint output

1000< Weight<2000 display OK

Weight>2000 display HI

11 Press【﹡】

〖 0〗

Press to confirm configurations and go to calibration

menu

Note:one time accumulation is allowed for weighing once. Following accumulation is

allowed for weighing only when displayed value is below 20 d.

☆ Optional LO

—OK—HI and two setpoint output PCB

☆ Press

【→】key for two seconds, LCD display for LED are turned on(A1GB-L)

7 Calibration

Calibration should be done after setting the parameters

Step

Operation

Displaying

Comments

1

Press【﹡】

Press【→】

〖CAL SP〗

〖CAL 00〗

Enters calibration

Zero the scale

2

Press【﹡】

〖------〗

〖 3000〗

Starts zero calibration and wait for calibration to

complete

3

Load

the

standard

weight for F·S on the

platform press【﹡】

〖------〗

〖 3000〗

Starts calibration and wait for calibration to

complete