beautypg.com

2 address: adr = 99, 5 power, 6 configuration – DIGIWEIGH DWP-PJ-P User Manual

Page 8

background image


DW-PJ Pallet Jack Scale

- 7 -

4.4 Continuous ASCII RS-232 data output format

4.4.1 Automatic Output

4.4.1.1 Address: Adr = 00

The ASCII data format is

“=,X1,X2,

X3,X4,X5,X6

=,X1 , X2, X3 , X4, X5, X6
X1, X2, X3, X4, X5, X6 is weight data.
If the weight is 〖 100.00〗kg, the continuous output is = 00.001 = 00.001=

4.4.1.2 Address: Adr = 99

The ASCII data format is

“=,X6,X5,X4,X3,X2,X1”

=,X6 , X5, X4 , X3, X2, X1
X6, X5, X4, X3, X2, X1 is weight data..
If the weight is 〖 100.00〗kg, the continuous output is = 100.00 = 100.00=

4.4.1.3 Manual and Automatic Printing Output

Address: Adr = 01

—98

5 Power

In power off states, press 【ON】 key turn on the indicator . The indicator will check the
LED and display battery capacity〖bPt 82〗for 1.5 second.

6 Configuration

Connect load cells to the indicator and set following configuration parameters

Step

Operation

Displaying

Contents

1

Press

【→0←】

【ON】

Self test from

〖UEr 7.9〗〖0〗to

〖9〗〖 0〗

In power off states, press 【→0←】key and

【ON】 turn on power, segment check, and display
〖UE

—7.9〗edition No 1.5 second.

Press【﹡】

Press【﹡】

〖CAL SP〗

〖-SEt-〗

Enter of the scale

Enter the configuration setting modal

Press【→】

〖d 1〗

The number of scale divisions selected

3

Press【↑】

Press【↑】

〖d 2〗

〖d 0.1〗

0.001-0.002-0.005-10-20-50-100-200-500-0.10-0.20-

0.50-0.010-0.020-0.050-1-2-5-0.1-0.2-0.5-0.01-0.02-

0.05 For example:d = 0.1

4

Press【﹡】

Press【→】

Press【→】

Press【→】

Press【↑】

Press【↑】

〖 6000〗

〖000000〗

〖001000〗

Sets scale F·S

This manual is related to the following products: