beautypg.com

DIGIWEIGH DWP-800 DWP-102E Indicator User Manual

Page 8

background image

- 7 -

SECTION 5 RS-232 OUTPUT

Specifications: RS-232 output of weighing data, 1200/2400/4800/9600/19200 bps,1 stop bit, 8 data bits, No Parity.

Mode 1: Continue output

ST,GS 0.119,kg

ST for stable; US for unstable ; GS for gross weight; NT for net weight ; 0.119 for weighing value; kg for unit (kg, lb); HEX 0d and HEX oa : HEX

for hex; 0d for enter; 0a for next line.

Mode 2: Automatic output

Automatic output mode is same as continue output. The only difference is:

After weighing, when take away the goods display returns to zero, then transfers.

Data Format for normal weighing operations. Examples as follow:

Mode 3: Print. Setting Lab2

SECTION 6 ERROR CODES

ERROR

CODES

DESCRIPTION

RESOLUTION

- - OL - -

Over range

Remove weight from the scale.

Err 1/Err 2

Date and Time not right

Reset again.

Err 4

Zero Setting Error

The scale was outside the normal zero setting

range either when it was turned on or when

the ZERO Key was pressed. Remove weight

from the scale and try again. Use the TARE

Key to set the display to zero value.

DATE 2000/02/11

TIME 21:26:56

NO. 001

TARE 0.000kg

NET 0.076kg

GROSS 0.076kg

DATE 2000/02/11

TIME 21:27:41

NO. 002

TARE 0.125kg

NET 0.035kg

GROSS 0.160kg

----------------

NO. 002

TOTAL 0.111kg