SHif

PV shift value

(offset).

Moves the display temperature to the oven

temperature

0

Filt

PV filter.

If the process value is unstable to read, increasing this

value steadies the input signal.

0

out1

Output 1

function.

Function of the output. Must be set to Hi. for High Limit

control.

Hi.

o1.Hy

Output 1

hysteresis

value.

Amount of degrees the temperature must be below the

setpoint temperature before the High Limit can be reset 2.0

HSP.L

Lower limit of

HSP1.

Minimum temperature setting for the High Limit

0 (32 if

F)

HSP.H

Upper limit of

HSP1.

Maximum temperature setting for the High Limit

204 (400 if

F)

out2

Second output

Despatch does not use this function

None

diSP

Normal display

format.

Used to select the display in normal condition:

PV = Process Value

SP! = High Limit Setpoint

SAFE = The word “safe” in normal condition

PV

PuHi

Process value

high

Read-Only Data: Maximum process value since last

UNLOCK operation.

None

PuLo

Process value

low

Read-Only Data: Minimum process value since last

UNLOCK operation.

None

t.Abn

Abnormal time

Total accumulated minutes of abnormal conditions

since last UNLOCK operation.

None

5.4.8.

Change High Limit instrument Display from Centigrade to

Fahrenheit

Warning!

Changing program parameters alters the function of the Control.

Proceed carefully and fully understand each parameter before

changing that parameter.

Configure the High Limit instrument for either °C or °F. Use the following steps to change High

Limit instrument from displaying °C to °F (and for changing back):

1. If

on the High Limit instrument is lit, press

and hold for four (4) seconds to

enable and .

2. Press and hold

for four (4) seconds to enter Setup mode.

3. Press

until

unit

displays.

4. Press or to display value.

5. Use or to move to the desired setting.
6. Press

to enter the value and advance to the next parameter.