Ct

Cycle Time Expressed in seconds, value shows total time for one

ON/OFF cycle of the controller output during the

proportional action

1

FL

Input Filter

Expressed in seconds. Use this function when the PV may

fluctuate greatly (for example, when an input signal

contains noise)

0

bS

PV Bias

(Offset)

Expressed in degrees, from –199 to 999. Use this

parameter to set the actual oven temperature to the

controller display.

0

LoC

Key Lock

Provides levels of access to the controller:

0 = No key lock, full access to controller

1 = Prevents changing of all parameters except

setpoint

2 = Prevents all parameters from being changed

including the setpoint

-1 = Set to enter the Set-up parameter setting display.

1

nr

Manual

Reset

Only when I & d are OFF. Expressed in percent.

Controller outputs this value when process variable equals

setpoint, but only if I = 0.

N/A

HyS

Hysteresis

Only when CtL is OFF, change CtL to PID)

Expressed in degrees. When CtL = OFF, this value

determines the change in temperature needed to turn

controller output from full OFF to full ON.

N/A

5.4.2.2. Control Instrument Operating Mode Notes

The controller is factory set to

LoC

=1 (allowing only the setpoint to be changed). This setting

prevents inadvertent changing of control parameters.

If necessary to alter control parameters, change the

LoC

=1 to

LoC

=0. See Section 5.4.2.1 for

more information.

When

LoC

= -1, parameters are displayed in the order shown in the Set-Up Parameters table

(Table 5).

Note: If unable to change setup parameters:

o

Change the

LoC

= 0

o

Press the

until

LoC

appears again

o

Change

LoC

= -1

o

Press the

Table 5. Control Instrument Set-up Parameters.
Code Name

Description

Settings

In

Input Type

Set for type of input and whether PV is in °C or °F

5
(35 if

F)

SPH

Setpoint High

The maximum setpoint limit for oven. The user cannot

set the setpoint above the maximum setpoint.

204 (400 if

F)

SPL

Setpoint Low

The minimum setpoint limit of oven. The user cannot

change the setpoint below this lower setpoint limit.

0 (32 if

F)