5.4.2.

Control Instrument Parameter Programming Mode

Warning!

Changing program parameters alters the function of the Control.

Proceed carefully and fully understand each parameter before

changing that parameter.

Control instrument parameters are set through the Operating and Set-up modes. In most

applications, it is not necessary to alter the oven settings. The following instructions describe how

to access, view and if desired, change the parameters. Once the Operating and Set-up modes are

accessed,

SP

will start blinking on and off. Table 4 and Table 5 explain the Operating and Set-up

Mode parameters.

The Control Instrument will not allow the display to be altered improperly. The Control

Instrument will automatically exit the Parameter Programming mode if no keys are pressed for

about two (2) minutes.

5.4.2.1. Entering Control Instrument Operating and Set-up Mode

To enter the Operating and Set-up Mode (Figure 4):

1. Press

for three (3) seconds.

2. Press

until the desired parameter displays. See Table 4 for more information.

3. Press ▲ or ▼ to display the value.

4. Use ▲ or ▼ to move to the desired setting.

5. Press

to enter the value.

6. Press

and hold for three (3) seconds to return to the display mode.

Table 4. Control Instrument Operating Parameters.
Code Name

Description

Settings

CtL

Control

Mode

Determines whether controller functions as a time

proportional or an on/off control.

PID

At

Auto-tuning OFF for PID tuning, ON for controller to tune process

OFF

P

Proportional

Band

Expressed in degrees, value determines the band width

on both sides of the setpoint within which the control

provides proportional control.

4

(8 if

F)

4

I

Integral

Time

Expressed in seconds, value corrects for errors in actual

temperature versus the setpoint

15

d

Derivative

Time

Expressed in seconds, value shows the effect of the

derivative time is in direct proportion to the time setting

0

4

If P is not displayed the Control Mode (CtL) must be first set to PID.