• Pressing
the
↓ key will next display ‘→Calibration’ on Line 2.
Press the ‘ENTER’ key and ‘
→C0’ will appear on Line 3 and a
numerical value on Line 4. Press the ‘ENTER’ key and use the
arrow keys to enter the calibration factors that will accompany the
sensor. Once the ‘C0’ value is entered press the ‘ENTER’ key to
store that value to memory. Line 3 will show ‘
→C0’; press the ↓
key to move to ‘
→C1’, press ‘ENTER’, and use the arrow keys to
enter ‘C1’ values. Once the values are entered, press ‘ENTER’
and press the
↓ key to move to ‘→C2’ and enter those values and
press ‘ENTER’ and the ‘BACK’ key.
• The display will show ‘→Calibration’ on Line 2. Press the ↓ key to
scroll to ‘
→Flow Direction’ on Line 2. If it is necessary to change
the flow direction (if the sensor was installed with the Flow
Direction arrow on the band opposite the actual flow direction
within the pipe or the process flow has changed direction) press
the ‘ENTER’ key and toggle the
↑ or ↓ to change flow direction.
Once changed press the ‘ENTER’ key.
•
The display will show ‘
→Flow Direction’ on Line 2. Press the ↓
key to scroll to ‘OP Mode’. If ‘Flow/GVF/SOS’ is not displayed,
press the ‘ENTER’ key and then scroll using the
↓ key to select
‘Flow/GVF/SOS’. Then press the ‘ENTER’ key.
• Press
the
↓ key to scroll to ‘→Set Date/Time’ on Line 2. Press the
‘ENTER’ key and the current saved date and time will be displayed
on Line 4 of the display. Use the arrow keys (
↑ and ↓ to scroll
through the values and
← and → to change character) to enter the
date and time. Note: Time is in 24 hour format. Once the date
and time have been set press the ‘ENTER’ key.
• Press
the
↓ key to scroll to ‘→Set Date Format’ on Line 2. Press
the ‘ENTER’ key and the current date format will be displayed on
Line 4 of the display. Use the arrow keys (
↑ and ↓) to scroll to the
desired format. Press the ‘ENTER’ key to set the desired date
format followed by the ‘BACK’ key.
• At this point ‘Basic Config’ is displayed on Line 1 of the display.
The user can re-enter this menu if desired by pressing the
‘ENTER’ key or by pressing the
↑ or ↓ arrows to move to other
Level 1 menus.
•
Press the ‘EXIT’ key to return to ‘Operational Mode’.
Entry of inputs to the other Level 1 menus is by the same process as
used in ‘Basic Config’.
Copyright © 2007 CiDRA Corporation
Page 3-12
20984-01 Rev 01