Check / set gain, Test gain – CiDRA SONARtrac Quick Start User Manual
Page 12

3.2.1.2.2 Check / Set Gain
Check / Set Gain allows the user to check the current gain setting in
the pre-amplifier and to manually set it to one of four settings 1, 4.65,
21.55, and 98.65.
•
Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.
•
The ‘Basic Config’ menu is displayed on Line 1 of the display.
•
Press the
↓ key to scroll to the ‘Diagnostics’ menu on Line 1.
Press the ‘ENTER’ key to access the options available under that
menu.
•
‘Sensor Check’ will appear on Line 2 of the display.
•
Press the
↓ key to scroll to ‘Gain’ on Line 2 of the display.
•
Press the ‘ENTER’ key to access the options available under that
menu.
•
‘Autoset Gain’ will appear on Line 3 of the display.
•
Press the
↓ key to scroll to ‘Check/Set Gain’ on Line 3 of the
display.
•
Press the ‘ENTER’ key and ‘Gain=XXXX Set=YYYY’ will appear
on Line 4 of the display.
•
Press the ‘ENTER’ key and the ‘Set’ value will be highlighted.
•
Press the
↓ key to scroll to a new ‘Set’ value.
•
Press the ‘ENTER’ key and the ‘Set’ value will be entered in the
pre-amplifier.
•
Press the ‘EXIT’ key to return to ‘Operational Mode’.
3.2.1.2.3 Test
Gain
‘Test Gain’ performs a test of the system electronics to determine if
the gain setting is optimal. Test results inform the user if the gain is
correct, or, if there is too much or too little gain; however, it does not
reset the gain. Resetting gain must be done using the ‘Autoset Gain’
or ‘Check / Set Gain’ commands. ‘Test Gain’ should be run while the
process is operating at normal conditions to avoid getting an incorrect
gain value.
•
Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.
•
The ‘Basic Config’ menu is displayed on Line 1 of the display.
•
Press the
↓ key to scroll to the ‘Diagnostics’ menu on Line 1.
Press the ‘ENTER’ key to access the options available under that
menu.
•
‘Sensor Check’ will appear on Line 2 of the display.
•
Press the
↓ key to scroll to ‘Gain’ on Line 2 of the display.
Copyright © 2007 CiDRA Corporation
Page 3-4
20984-01 Rev 01