Vf/gvf-100 system initial setup, Vf/gvf-100 system initial setup -9 – CiDRA SONARtrac Quick Start User Manual
Page 17

desired format. Press the ‘ENTER’ key to set the desired date
format followed by the ‘BACK’ key.
• At this point ‘Basic Config’ is displayed on Line 1 of the display.
The user can re-enter this menu if desired by pressing the ‘ENTER’
key or by pressing the
↑ or ↓ arrows to move to other Level 1
menus.
•
Press the ‘EXIT’ key to return to ‘Operational Mode’.
Entry of inputs to the other Level 1 menus is by the same process as
used in ‘Basic Config’.
3.3.2
VF/GVF-100 System Initial Setup
In order to conduct measurements, the ‘Basic Config’ setup must be
completed. If power goes off, this setup will remain in memory and
does not have to be re-entered.
• Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.
• When
‘Basic Config’ menu is displayed on Line 1 of the display
press the ‘ENTER’ key to enter the options available under that
menu.
• ‘→Sensor Serial #’ will appear on Line 2 of the display. Press the
‘ENTER’ key and the current sensor band serial number entered in
the transmitter will be displayed on Line 4 of the display. The user
can then use the arrow keys (
↑ and ↓ to scroll through the values
and
← and → to change position) to enter the sensor serial
number. Once all digits are entered, press the ‘ENTER’ key to
save to memory. ‘
→Sensor Serial #’ will again be displayed on
Line 2.
• Next, press the ↓ key to scroll to the ‘Pipe Size’ menu on Line 2.
Press the ‘ENTER’ key to access the options available under that
menu. Note: it is necessary to select only one of the following
options.
• The first choice on the ‘Pipe Size’ menu is ‘→ID / Wall’ shown on
Line 3 of the display. If this value is known it can be entered here.
Press the ‘ENTER’ key and the current saved inner diameter and
wall thickness will be shown on Line 4 of the display. Use the
arrow keys (
↑ and ↓ to scroll through the values and ← and → to
change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
‘ENTER’ key to save the value to memory.
• The second choice on the ‘Pipe Size’ menu is ‘Size/Sched’ (pipe
Size / Schedule). If this value is known press the ‘ENTER’ key.
Use the arrow keys to enter the values and then press the
‘ENTER’ key to save to memory.
Copyright © 2007 CiDRA Corporation
Page 3-9
20984-01 Rev 01