Initial diagnostic checks, Process on-line, Sensor test – CiDRA SONARtrac Quick Start User Manual
Page 10: Initial diagnostic checks -2, Process on-line -2, Table 1, Keypad functions in operational and menu modes -2, 2 initial diagnostic checks

Menu Mode
Key
Operational
Mode
Navigation Editing
Dialog
Box
Up
Enter Menu
Mode
Cycle Menu Item
Change current value at
cursor position
Exit Dialog
Down
Enter Menu
Mode
Cycle Menu Item
Change current value at
cursor position
Exit Dialog
Left
Enter Menu
Mode
NA
Change cursor position
Exit Dialog
Right
Enter Menu
Mode
NA
Change cursor position
Exit Dialog
EXIT
Clears Alarm
Exit Menu
Exit Menu
Exit Dialog
BACK
Enter Menu
Mode
Exit Menu from Main
Menu or back up one
level in menu tree
Exit editing mode
without saving
Exit Dialog
ENTER
Enter Menu
Mode
Change menu level or
start editing
Exit editing mode and
save current value
Exit Dialog
Table 1
Keypad Functions in Operational and Menu Modes
3.2
Initial Diagnostic Checks
3.2.1
Process On-Line
The SONARtrac meter can be installed and configured with the
process in operation. Ideally, configuration is made with the process
operating at normal flow rates and operating conditions.
If the process is on-line (flowing) the following system checks should
be made.
3.2.1.1 Sensor
Test
A Sensor Test all ‘ON’ sensors. (Individual sensors may be turned off
in the ‘Customize>Sensor Setup>State’ menu. Sensors should never
be turned off unless directed to do so by Technical Support
Personnel.) The test will take several seconds to perform. After the
test is completed, the screen will display test results.
A Sensor Test is made as follows:
•
Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.
•
The ‘Basic Config’ menu is displayed on Line 1 of the display.
•
Press the
↓ key to scroll to the ‘Diagnostics’ menu on Line 1.
Press the ‘ENTER’ key to access the options available under that
menu.
•
‘Sensor Check’ will appear on Line 2 of the display.
Copyright © 2007 CiDRA Corporation
Page 3-2
20984-01 Rev 01