Operating Menu Setup

3.3.1

VF-100 System Initial Setup

In order to conduct measurements, the ‘Basic Config’ setup must be
completed. If power goes off, this setup will remain in memory and
does not have to be re-entered.

Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.

• When

Basic Config’ menu is displayed on Line 1 of the display

press the ‘ENTER’ key to enter the options available under that
menu.

Sensor Serial #’ will appear on Line 2 of the display. Press the

ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys (

and to scroll through the values and

and

to change position) to enter the sensor band serial number

found on the sticker attached to the sensor band. Once all digits
are entered, press the ‘ENTER’ key to save to memory. ‘

→Sensor

Serial #’ will again be displayed on Line 2.

• Next, press the key to scroll to the ‘Pipe Size’ menu on Line 2.

Press the ‘ENTER’ key to access the options available under that
menu. Note: it is necessary to select only one of the following
options.

• The first choice on the ‘Pipe Size’ menu is ‘→ID / Wall’ shown on

Line 3 of the display. If this value is known it can be entered here.
Press the ‘ENTER’ key and the current saved inner diameter and
wall thickness will be shown on Line 4 of the display. Use the
arrow keys (

and to scroll through the values and and to

change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
ENTER’ key to save the value to memory.

• The second choice on the ‘Pipe Size’ menu is ‘Size/Sched’ (pipe

Size / Schedule). If this value is known, press the ‘ENTER’ key.
Use the arrow keys to enter the values and then press the ‘ENTER
key to save to memory.

• The third choice on the ‘Pipe Size’ menu is ‘OD / Wall’ displayed

on Line 3. Use the arrow keys (

and to scroll through the values

and

and to change character) to enter the pipe outer

diameter. Units can be inches or millimeters. Once all digits are
entered, press the ‘ENTER’ key to save the value to memory.
Note: Once a pipe size has been entered, re-entering the ‘Pipe
Size
’ menu and selecting ‘Size/Sched’ may cause errors in scaled
output values.

Copyright © 2007 CiDRA Corporation

Page 3-7