beautypg.com

2 3 1 front and rear panels – C.B.S. Scientific DGGEK-4801 User Manual

Page 5

background image

8

9

www.cbsscientific.com

Cipher DGGE Instructions 2/28/14







Front View

Rear View

1. Display

11.

Pump Speed Switch

2. Preset Temperature Buttons

12.

Identification Label

3. Safety Set Reset Button

13.

Pump Inlet

2

4. Safety Set Indicator Knob

14.

Reservoir Purge

2

5. Cooling Light

1

15.

Pump Outlet

2

6. Power ON / OFF Button

16.

AC Input

7. Heating Light

17.

Circuit Breaker / AC Power Switch

8. °F Light

9. Select/Set Knob (Push to set, turn to adjust)
10. °C Light

1

Present on Refrigerating/Heating models only

2

Built in on reservoir models only

1

2

3

4

6

5

7

9

8

10

11

12

13

17

14

16

15

2 3 Operating Heater/Stirrer/By-Pass Pump (Controller)

1) Attach power cord to heater/stirrer and plug into wall outlet.

(comes factory-mounted onto DGGE tank)

2 3 1 Front and Rear Panels





1. Temperature Sensor
2. Pump Shaft and Impeller
3. Pump Outlet Nozzle

1

4. Over-Temperature Sensor
5. Heater Coils

1

Rotates on Immersion Circulators







Temperature Stability

±0.05°C

Readout Accuracy

±0.5°C

Over-Temperature Protection

Yes, user-adjustable

Low-Liquid Protection

Yes

Heater

1100W – 115V,

1600 W – 240V

Pump Speeds

2-speed, 9 liters per minute or 15 liters per minute

Environmental Conditions:
 Indoor Use Only

 Over Voltage: Category ll

 Maximum Altitude: 2000 meters

 Operating Ambient: 5° to 30°C

 Relative Humidity: 80% for temperatures to 30°C

 Pollution Degree: 2

 Class 1: Residential, Commercial, Light Industrial

 Class 2: Heavy Industrial







Locate the Circulator on a level surface, free from drafts and out of direct sunlight. Do not place it

where there are corrosive fumes, excessive moisture, high room temperatures, or excessive dust is
present.

Refrigerating/Heating Circulators must be a minimum of four inches (102mm) away from walls or
vertical surfaces so air flow around the unit is not restricted.

To help prevent voltage drops, position the Circulator as close as possible to the power distribution

panel and a properly grounded outlet. The use of an extension cord is not recommended.

Warning: These units are equipped with over-temperature protection (Safety Set). A low liquid level

or failure to set the Safety Set and properly immerse the heater may result in heater burnout and triac
failure. While operating, do not allow the heater to contact any potentially flammable materials, such

as plastic racks or sides of plastic tanks, as a fire hazard may result.

5

3

1

2

4

2 3 2

Controller Assembly

2 3 3

Controller Specifications

1. Temperature Sensor
2. Pump Shaft and Impeller
3. Buffer By-Pass Tubing Adapter
4. Over-Temperature Sensor
5. Heater Coils

1.

5.

4.

3.

2.

This manual is related to the following products: