DCR-1006A Ultrasonic Controller

Rev. A1, 2/14

32

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

4-20mA Settings

NOTE: The 4-20mA Set Points must be entered in the units being displayed.
For example if the display is setup to read in gallons, then the 4mA and 20mA
Set Points would be set in gallons.

SET4MA (4mA Set Point):

Sets the reading to be associated with a 4mA output.

SET20M (20mA Set Point):

Sets the reading to be associated with a 20mA output.

CAL4MA (4mA Calibration):

Used to adjust the output current at the low end of the analog range (4mA

endpoint). Increasing/decreasing the parameter value will increase/decrease

the output current.

CAL20M (20mA Calibration):

Used to adjust the output current at the high end of the analog range (20mA

endpoint). Increasing/decreasing the parameter value will increase/decrease

the output current.