8 gateway preferences icon, 1 general tab, 2 info tab – Eppendorf TCA-3 Temperature Monitoring System User Manual
Page 24: Gateway preferences icon 4.8.1, General tab, Info tab

Operation
TCA-3 Temperature Monitoring System
English (EN)
24
4.8
Gateway preferences icon
4.8.1
General tab
The General tab allows the users to name the TCA-3 monitoring system (TCA-3 monitoring systems will be
listed on the screen in alphabetical order), set the sampling rates (how often measurements are made) and
record notes about the TCA-3 monitoring system.
The following sampling rates can be selected:
•
Store data every: This option records data if the measurements are within limits which can be slower
•
Check for alarm every: This option checks measurements against alarm limits which can be faster
4.8.2
Info tab
The Info tab provides information about the TCA-3 monitoring system and firmware. This information can
be useful when diagnosing system problems remotely.