Eppendorf Universal Adapter UVette User Manual
Universal adapter uvette® — instructions for use, Universal adapter uvette, Select a wavelength of 550 nm
Universal Adapter UVette® — Instructions for Use
Instructions for Use
\ \m\
Universal-Adapter UVette®
See
on page
Fig.
Tab.
S.
Leaflet
1
Universal Adapter UVette
®
To enable the Universal Adapter to be used with UVette
®
in as many spectro-
photometers as possible, it is first necessary to determine the light center height of the
spectrophotometer.
If no specifications for the light center height are available from the
manufacturer, it is possible to determine this height using the
measuring gauge:
•
Select a wavelength of 550 nm.
•
Insert the measuring gauge vertically into the cuvette holder (in its
original condition) until it reaches the bottom.
•
Read the height of the light spot.
The most common light center height is 15 mm, which enables the
adapter to be used without being altered.