Eppendorf epMotion Dispensing tools User Manual
1main illustration 2 product description, 3delivery package, 4operation
Instructions for use
epMotion
®
Dispensing tools
English (EN)
Instructions for use
epMotion® Dispensing tools
see
on p.
Fig.
Tab.
p.
English (EN)
Instructions for use
1
Main illustration
2
Product description
Dispensing tools for epMotion dispensing systems are piston-stroke pipettes which
operate according to the air cushion principle. If the piston in the dispensing tool moves
upwards the liquid is aspirated into the pipette tip. If the piston in the dispensing tool
moves downwards, the liquid is dispensed from the pipette tip. In multi-channel
dispensing tools the pistons move simultaneously in all channels.
3
Delivery package
• Instructions for use
• Eppendorf certificate
For multi-channel dispensing tools the following are also enclosed:
• Tool for the removal of worn O-rings
• Mounting aid for O-rings
• 8 replacement O-rings
4
Operation
Use the dispensing tools in accordance with the descriptions in the corresponding
epMotion operating manual.
5
Cleaning
Cleaning single-channel dispensing tools
1. Remove ejector for cleaning.
2. Wipe off contaminations on the tip cones and the surfaces with a damp cloth.
Cleaning multi-channel dispensing tools
Wipe off contaminations on the tip cones and the surfaces with a damp cloth.
Replace o-rings on the tip cones once per year or more often if required.
TS 50
TS 300
TS 1000
TM 50-8
TM 300-8
TM 1000-8
NOTICE! Any contamination of the gold contacts will have an impact on the
performance of the dispensing tools.
Do not touch the gold contacts.
50
50
300
300
1000
1000