Eppendorf Combitips advanced User Manual
Page 20

Technical data
Combitips advanced
®
English (EN)
20
6.4
Errors with the Multipette (X)stream/Repeater (X)stream
Combitip advanced
Volume range
Testing
volume
Error limits
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.1
mL
white
Increment:
0.1
μL
1
μL
–
100
μL
10
μL
±1.6 ±0.16
±2.5 ±0.25
50
μL
±1.0 ±0.5 ±1.5 ±0.75
100
μL
±1.0 ±1.0 ±0.5 ±0.5
0.2
mL
light blue
Increment:
0.2
μL
2
μL
–
200
μL
20
μL
±1.3 ±0.26
±1.5 ±0.3
100
μL
±1.0 ±1.0 ±1.0 ±1.0
200
μL
±1.0 ±2.0 ±0.5 ±1.0
0.5
mL
purple
Increment:
0.5
μL
5
μL
–
500
μL
50
μL
±0.9 ±0.45
±0.8 ±0.4
250
μL
±0.9 ±2.25
±0.5 ±1.25
500
μL
±0.9 ±4.5 ±0.3 ±1.5
1
mL
yellow
Increment:
1
μL
10
μL
–
1
000
μL
100
μL
±0.9 ±0.9 ±0.55
±0.55
500
μL
±0.6 ±3.0 ±0.3 ±1.5
1 000
μL
±0.6 ±6.0 ±0.2 ±2.00
2.5
mL
green
Increment:
2.5
μL
25
μL
–
2
500
μL
250
μL
±0.8 ±2.0 ±0.45
±1.125
1 250
μL
±0.5 ±6.25
±0.3 ±3.75
2 500
μL
±0.5 ±12.5 ±0.15 ±3.75
5
mL
blue
Increment:
5
μL
50
μL
–
5
000
μL
500
μL
±0.8 ±4.0 ±0.35
±1.75
2 500
μL
±0.5 ±12.5 ±0.25 ±6.25
5
000
μL ±0.5
±25
±0.15
±7.50
mL
mL
mL
±
%
±
mL
±
%
±
mL
10
mL
orange
Increment:
0.01
mL
0.1
mL
–
10
mL
1
mL
±0.5
±0.005
±0.25
±0.0025
5
mL
±0.4
±0.02
±0.25
±0.0125
10
mL
±0.4
±0.04
±0.15
±0.015
25
mL
red
Increment:
0.025
mL
0.25
mL
–
25
mL
2.5
mL
±0.3
±0.0075 ±0.35
±0.0088
12.5
mL
±0.3
±0.0375
±0.25
±0.0313
25
mL
±0.3
±0.075
±0.15
±0.0375
50
mL
light gray
Increment:
0.05
mL
0.5
mL
–
50
mL
5
mL
±0.3
±0.015
±0.5
±0.025
25
mL
±0.3
±0.075
±0.20
±0.05
50
mL
±0.3
±0.15
±0.15
±0.075