3 errors with the multipette plus/repeater plus, Errors with the multipette plus/repeater plus – Eppendorf Combitips advanced User Manual
Page 19

Technical data
Combitips advanced
®
English (EN)
19
6.3
Errors with the Multipette plus/Repeater plus
Test conditions and test evaluation in compliance with ISO
8655,
Part
6. Analytical balance with
evaporation protection inspected by the office of weights and measures.
• Number of determinations: 10
• Use of water in accordance with ISO
3696
• Inspection at 20
°C – 25
°C
±0.5
°C
• Dispensing on the inner wall of the tube
Combitip advanced
Volume range
Error limits
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.1
mL
white
2
μL
±1.6
±0.032
±3.0
±0.06
20
μL
±1.0 ±0.2 ±2.0
±0.4
0.2
mL
light blue
4
μL
±1.3
±0.052
±2.0
±0.08
40
μL
±0.8
±0.32
±1.5
±0.6
0.5
mL
purple
10
μL
±0.9
±0.09
±1.5
±0.15
100
μL
±0.8 ±0.8 ±0.6
±0.6
1
mL
yellow
20
μL
±0.9
±0.18
±0.9
±0.18
200
μL
±0.6 ±1.2 ±0.4
±0.8
2.5
mL
green
50
μL
±0.8 ±0.4 ±0.8
±0.4
500
μL
±0.5 ±2.5 ±0.3
±1.5
5
mL
blue
100
μL
±0.6 ±0.6 ±0.6
±0.6
1
000
μL
±0.5
±5.0
±0.25
±2.5
10
mL
orange
200
μL
±0.5
±1.0
±0.6
±1.2
2
000
μL
±0.5
±10
±0.25
±5.0
25
mL
red
500
μL
±0.4
±2.0
±0.6
±3.0
5
000
μL
±0.3
±15
±0.25
±12.5
50
mL
light gray
1
000
μL
±0.3
±3.0
±0.5
±5.0
10
000
μL
±0.3
±30
±0.3
±30
The test volumes for the systematic and random errors of the Multipette (X)stream/Repeater
(X)stream comply with the requirements of ISO 8655, part 5.