2 errors with the multipette m4/repeater m4, Errors with the multipette m4/repeater m4 – Eppendorf Combitips advanced User Manual
Page 17

Technical data
Combitips advanced
®
English (EN)
17
6.2
Errors with the Multipette M4/Repeater M4
Combitip advanced
Testing volume
Error limits
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.1
mL
white
Increment:
1
μL
2
μL
±1.6
±0.032
±3.0
±0.06
20
μL
±1.0 ±0.2 ±2.0
±0.4
0.2
mL
light blue
Increment:
2
μL
4
μL
±1.3
±0.052
±2.0
±0.08
40
μL
±0.8
±0.32
±1.5
±0.6
0.5
mL
purple
Increment:
5
μL
10 μL
±0.9
±0.09
±1.5
±0.15
100
μL
±0.8 ±0.8 ±0.6
±0.6
1
mL
yellow
Increment:
10 μL
20
μL
±0.9
±0.18
±0.9
±0.18
200
μL
±0.6 ±1.2 ±0.4
±0.8
2.5
mL
green
Increment:
25
μL
50
μL
±0.8 ±0.4 ±0.8
±0.4
500
μL
±0.5 ±2.5 ±0.3
±1.5
5
mL
blue
Increment:
50
μL
100
μL ±0.6 ±0.6 ±0.6
±0.6
1
000
μL
±0.5
±5.0
±0.25
±2.5
10
mL
orange
Increment:
0.1
mL
200
μL
0.2
mL
±0.5 ±1.0
±0.6
±1.2
2
000
μL
2
mL
±0.5 ±10
±0.25
±5.0
25
mL
red
Increment:
0.25
mL
500
μL
0.5
mL
±0.4 ±2.0
±0.6
±3.0
5
000
μL
5
mL
±0.3 ±15
±0.25
±12.5
50
mL
light gray
Increment:
0.5
mL
1
000
μL
1
mL
±0.3 ±3.0
±0.5
±5.0
10
000
μL
10
mL
±0.3 ±30
±0.3
±30