8 maintenance, 1 routine maintenance, 2 opening the service compartment – Eppendorf New Brunswick Excella E-24 Shaker User Manual
Page 29: 1 fuse replacement, Maintenance 8.1, Routine maintenance, Opening the service compartment 8.2.1, Fuse replacement, Replace (see, On p. 29)
Maintenance
New Brunswick
™
Excella
®
E-24 Shaker
English (EN)
29
8
Maintenance
8.1
Routine maintenance
No routine maintenance schedule is required for the E-24.
Clean the incubator shaker occasionally using a cloth with conventional household (non-abrasive) cleaner.
To ensure proper air flow in and around the incubator shaker, vacuum or sweep the area around the
incubator shaker to remove dust and other debris.
8.2
Opening the service compartment
The service compartment contains the shaker’s electronic and temperature control components. Normally,
this compartment should be accessed by authorized service technicians only. You may, from time to time,
need to remove the access panel in order to replace fuses.
8.2.1
Fuse replacement
The user can replace one fuse on the E-24, located behind the front bezel panel (on the PC board). To
access the fuse:
1. Turn the power off and unplug the shaker.
2. Open the front cover.
3. Remove the 4 fastners that hold the front panel in place.
Set them aside for the reuse.
4. The fuse is located on the PC board located on the right side of the base weldment.
5. Access the fuse by using a coin or a blade screwdriver to turn and release the spring-loaded cap.
6. Replace it with a new one of the same type and rating.
WARNING!
Before opening the service compartment, always turn off the shaker and disconnect the
power cord from the power supply.
WARNING!
When replacing fuses, always turn off the shaker and disconnect the power cord from the
power supply.
Fuse holder number
Function
Type
Rating
F1
Heater
Slo Blo 8 A
8.0 A