7 troubleshooting, Troubleshooting, 7troubleshooting – Eppendorf New Brunswick Excella E-24 Shaker User Manual
Page 27
Troubleshooting
New Brunswick
™
Excella
®
E-24 Shaker
English (EN)
27
7
Troubleshooting
Symptoms
Possible causes
Solutions
Shaker does not run
• Power cord is not plugged in
and/or power switch is off
Plug in power cord to working
electrical outlet
Turn on the power switch
• Lid is open,
Lid
is indicated on
the display
Open and reclose the lid firmly
• Recently replaced fuse may
not be seated properly
Reinstall the fuse carefully
• On/off switch is not working
• Defective main board
• Defective display controller
board
• Jammed shaking mechanism
• Defective motor
• Drive belt out of alignment or
worn
Call for service
Shaker runs slowly and/or no
speed indication
• Incorrect speed calibration
Recalibrate shaking speed (see
Speed calibration on p. 20)
• Defective main board
• Defective motor
• Drive belt out of alignment or
worn
Call for service
Shaker does not run at set speed
• Shaker is overloaded and/or
you are using baffled flasks
Remove some contents and
balance load
• Defective motor
• Drive belt is out of alignment
or worn
Call for service
Operating noise
• Load out of balance
Unload all contents and reload
• Loose component(s) in
platform, subplatform and/or
drive assembly
Call for service
Incubator does not reach set
temperature
• Heater fuse blown
• Ambient temperature too high
or too low
Cool or heat the room as
needed
• Compressor over-pressure
switch activated
• Defective heater
Call for service
• Incorrect temperature
indication