4 installation, 1 selecting the location, 2 unpacking the device – Eppendorf New Brunswick Excella E-24 Shaker User Manual
Page 13: 3 utility requirements, Installation 4.1, Selecting the location, Unpacking the device, Utility requirements

Installation
New Brunswick
™
Excella
®
E-24 Shaker
English (EN)
13
4
Installation
4.1
Selecting the location
Select the location according to the following criteria:
• Suitable mains power connection according to the ID plate
• Stable, even and resonance-free base
• Well ventilated area and no direct sunlight to prevent additional temperature increases
• Ambient conditions of 10 °C to 35 °C, 20 % to 80 % non condensing
• Able to accommodate 200 lb
4.2
Unpacking the device
1. Remove the packing material.
2. Remove the transport securing device.
3. Use the details included in the scope of delivery to check that the delivery is complete.
4. Check all parts for damage in transit. Contact Eppendorf Service if parts are missing or transport
damage is present.
4.3
Utility requirements
The following utility requirements are needed for operation:
In all cases, voltage variations must not exceed ±10
%.
NOTICE! Damage due to overheating
Do not place the device close to sources of heat (e.g., radiator, drying cabinet).
Do not expose the device to direct sunlight.
Ensure there is adequate distance to the wall and adjacent devices, on all sides of the
device, in order to guarantee unobstructed air circulation.
Keep the packing material and transport securing device for later transport or storage.
Utility
Requirement
Electricity
• 230 V, 60 Hz, 1500 VA maximum